Operating Instructions | HERAcell® 150i / 240i
3
Installation of the device
3
Installation of the device
25
50115191 A / 08.2009
Mobile racks
Fastening on mobile racks:
If the devices are installed onto mobile racks, make sure that
the rollers [10] are secured with a locking brake during the
operation of the incubators and that the rollers are oriented
toward the front for increased stability.
Condensate formation during the operation of stacked devices:
If stacked devices of type HERAcell
®
150i are operated at an
ambient temperature of more than 26 °C, condensate formation
may occur on the upper device while the contra-con
decontamination routine is run for the lower device.
To prevent condensate formation, an adapter plate should be
placed between the two devices for thermal separation.
3.6 Retrofitting/Modifications
Retrofitting options
The following components can be retrofitted to the standard version:
HERAcell
®
150i
• 3-segment gas-tight screen (the gas-tight screen is installed in place of
the glass door),
• door hinge replacement on outer door and glass door (RH / LH door
opening),
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
HERAcell
®
240i
• 6-segment gas-tight screen (the gas-tight screen is installed in place of
the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
Modifications:
Retrofittings and modifications must only be performed by the
Technical Service of Thermo Electron LED.