Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
86
50115191 A / 08.2009
Setting low humidity:
If condensation occurs on the culture containers due
to high relative humidity, the humidity in the work
space can be set to a lower level. At the factory, the
device control is preset to high humidity (approx. 93%
relative humidity).
By enabling the Low Humidity feature, the relative
humidity in the work space is lowered from approx.
93% to approx. 90%. The modification requires an
extended adaption phase. To effectively prevent
condensation on culture containers, it must be used as
a permanent setting.
1. To toggle between two states [1]:
>
Press the
+
key [4].
or
>
Press the
-
key [2].
2. To accept and save the change:
>
Press the
E
NTER
key [3].
>
The system returns to the Options menu.
Low humidity:
The enabling/disabling of the Low
humidity function is entered into the event
list.
If the Low humidity function is enabled, the
corresponding icon [5] is displayed in the
main menu.
Fig. 53: Low humidity setting