Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
80
50115191 A / 08.2009
Setting the interface baud rate:
The input dialog allows the setting of the stepping rate
for data communication of the:
• RS232 interface,
• USB interface (optional).
The stepping rate of both interfaces can be changed
within the defined baud rates (9.600, 19.200, 38.400,
57.600 baud).
The standard stepping rate of both interfaces is 57.600
baud.
1. To increase the value:
>
Press the
+
key [4].
2. To reduce the value:
>
Press the
-
key [2].
3. The value change appears in the display [1].
4. To accept and save the change:
>
Press the
E
NTER
key [3].
>
The system returns to the Interface menu.
5. To activate the new settings:
>
Browse back to the main menu.
>
Wait for approx. 10 seconds and perform a
reset by switching the device off and on using
the power switch.
Fig. 45: RS 232 interface baud rate setting
Fig. 46: USB interface baud rate setting
Содержание Heracell 150i
Страница 156: ......