Operating Instructions | HERAcell® 150i / 240i
4
Description of the device
4
Description of the device
4
Description of the device
47
50115191 A / 08.2009
4.10.7 Shelf system
Fig. 16:
The support rails [1] of the shelf system are
perforated every 42 mm so that the support hooks [8]
can be inserted variably for any culture container size
required. The shelves [2] have an integrated tilt
protection and withdrawal stop. For details of the shelf
system, see Section „Start-up“.
4.10.8 Bottle turning device for HERAcell® 240i
(optional)
Fig. 16:
HERAcell
®
240i may be equipped with up to
four separate bottle turning devices. Each unit
consists of a roller insert [6] with a drive roller [4] and
four idle rollers [7] that are also rotated due to the
transmission function of the inserted cylinders [5].
Each unit is controlled separately as the drive roller is
connected to the device control unit by a connector [3]
at the right side of the device.
This modular layout also allows the operation of only
one bottle turning device in combination with several
shelves.
The arrangement of the idle rollers in the roller insert
frame depends on the configuration of the cylinders
used. The drive roller is always inserted into the right
roller guide. The position and the number of the idle
rollers in the frame then depend on the cylinder
diameter and on the number of cylinders. The figure
shows three examples of different idle roller positions
in dependence of the cylinder diameter.
The arrangement of the roller inserts in the work space
is designed alphabetically as a, b, c, and d with a being
the lowest and d being the uppermost roller insert.
The connections of the drive rollers are assigned
correspondingly to the device control unit. The drive
roller of the switching insert a must only be connected
to the lower receptacle a of the corresponding control
unit.
Connection assignment:
If the drive roller is not properly connected
to the control unit, the parameters cannot
be set correctly on the operating panel of
the device.
Fig. 16: Bottle turning device