13.5.1 Error memory data set structure scheme:
Therefore, the following information is transmitted in this data set:
• Created on 1 January 2006 at 15:55:34 hrs.
• A device error has occurred and the set temperature is too high.
13.5.2 Overview of the possible error messages in hex coding
Hex-coded error message
00 0b 01 06
0f 37 22
8000
0002
Faulty control loop type
0x00 = Error in temperature
control loop
00
Date / Day
0x0b = 11th day of the month
0b
Date / Month
0x01 = January
01
Date / Year
0x06 = 2006
06
Time / Hours
0x0f = 15 hrs.
0f
Time / Minutes
0x37 = 55 minutes
37
Time / Seconds
0x22 = 34 seconds
22
Device status
0x8000 = Device error
existing
8000
Error message
0x0002 = Value above set
value upper limit
0002
Hex code
Description / Type
0x00
Temperature control loop
0x01
CO
2
control loop
0x02
O
2
control loop
0x07
Water level
0x08
General device status