Operating Instructions | HERAcell® 150i / 240i
63
50115191 A / 08.2009
6
Operation
6.1 Preparing the device
The device must only be released for operation after all major measures for
the start-up have been taken (see Section 5, page 45).
Device check:
Device check
Prior to starting operation, the following device components must be
checked for their correct function
• Gas hoses must be seated tight on the connecting filter and must be
secured using a hose clamp.
• The access port must be capped.
• The pressure compensation opening must be permeable, its insert
must be installed in the work space.
• The glass door/gas-tight screen seal must not be damaged.
• The glass door measurement opening must be capped.
• The shelf system components must be installed safely.
• The gas humidification (optional) must be connected to the device-
integral gas supply and aligned parallel to the rear panel.
• The drive rollers of the bottle turning device (optional) must be properly
connected to the corresponding control unit.
• The idle rollers of the bottle turning device (optional) must be
positioned in accordance with the cylinder diameters.
Decontaminating the device work space:
Decontamination
Run the contra-con decontamination routine
(see chap. 9.4‚ pg. 104)
or
decontaminate the work space according to the hygiene regulations set
forth by the operator.
Hygiene regulations:
Prior to any operation, the user must clean and disinfect the
work space in accordance with the hygiene regulations set
forth by the operator to protect the cultures.
The “Principles of good microbiological proceedings” at the
annex of these instructions can be used as safety information
for personnel working with the device.
Water supply:
A sufficient quantity of treated water of the following quality
must always be available for the running operation:
• distilled, fully demineralized, and
• autoclaved, sterile.
When the filling level drops below a minimum value, it must be
possible to top up water during the work process.
Capacities:
• HERAcell
®
150i: 3.0 l
• HERAcell
®
240i: 4.5 l