Operating Instructions | HERAcell® 150i / 240i
9
Cleaning and disinfection
9
Cleaning and disinfection
103
50115191 A / 08.2009
of the measuring cell. The wheel and the cover can be autoclaved.
Removing blower wheel and cover:
Electric shock!
Contact with current-carrying components may cause a lethal
electric shock.
Prior to removing the blower wheel, switch the device off and
disconnect it from the power supply system.
1. Remove the two retaining screws of the cover using the supplied Allen
wrench (3 mm) and remove the cover.
2. The blower wheel is secured to the axle by a set screw. Remove the
set screw using the Allen wrench (2 mm) and pull the blower wheel
off.
Cleaning the work space and accessories:
1. Thoroughly remove dirt residues and deposits using a solution of
tepid water and dishwashing agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the water tray and wipe all surfaces
of the work space thoroughly dry.
4. Wipe accessories thoroughly dry.
Final disinfection:
Final disinfection
1. Spray disinfectant again onto the work space, the shelf system, and
the removed components and wipe them clean.
2. Allow the disinfectant to react as specified by the manufacturer.
3. Reinstall the shelf system and the removed components.
Functional check:
After the installation, check to see if the blower wheel is
securely attached to the axle and if it can rotate freely, then
secure cover using the screws.