Operating Instructions | HERAcell® 150i / 240i
13
Annex: Data communication
13
Annex: Data communication
50115191 A / 08.2009
131
Note on *3) disinfection and autostart run status:
Bit
Disinfection
auto-start
0x00
contra-con not active
auto-start not active
0x01
Initialization
Initialization
0x02
Await door opening time
Await door opening time
0x03
Await door closing
Await door closing
0x04
Start
Start
0x05
Heat
Heat
0x06
Hold
Perform counter-voltage
adjustment
0x07
Condensation
Waiting period 1
0x08
Cool-down
Set tolerance band
0x09
Dry
Establish stable humidity
0x0A
Await release
Perform counter-voltage
adjustment
0x0B
Stop
Waiting period 2
0x0C
-
Determine offset
0x0D
-
Read offset, check
0x0E
-
Release
0x0F
-
Stop