Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
85
50115191 A / 08.2009
Troubleshooting:
For a detailed overview of causes for errors
and their correction, please refer to the end
of this chapter!
7.11.3 Options
The input dialogs of the
O
PTIONS
category comprise all
settings of the functional device options:
• Alarm relay,
• Low humidity,
• Gas-tight screen,
• Water level sensor,
• Audible alarm,
• Bottle turning device (optional),
• O
2
(optional).
Setting the alarm relay:
The alarm relay is the interface for the connection of
the device-integral monitoring system to an external
monitoring system. Depending on the required input
signal of the external monitoring system, network
monitoring can be enabled or disabled. If network
monitoring is enabled, a power failure is detected as
an error. The relay cannot be switched off
(see chap.
5.10‚ pg. 61)
.
1. To toggle between two states [1]:
>
Press the
+
key [4].
or
>
Press the
-
key [2].
2. To accept and save the change:
>
Press the
E
NTER
key [3].
>
The system returns to the Options menu.
Fig. 52: Alarm relay setting