Operating Instructions | HERAcell® 150i / 240i
4
Description of the device
4
Description of the device
4
Description of the device
48
50115191 A / 08.2009
Receptacles:
To prevent corrosion due to moisture in the
work space, unassigned receptacles of the
bottle turning device must always be
plugged with the protective cap.
The rotating speed of the bottle turning devices is set
depending on two parameters:
• Diameter (of the cylinders used),
• drive roller set value.
The value is determined using a diagram
(see chap.
5.4‚ pg. 51)
.
4.10.9 Water pump
Fig. 17:
The water pump serves for suction cleaning of
the remainig water in the water reservoir. The water
drain works through gravity.
1. Place the water pump [1] on lowest shelf [5].
2. Feed the inlet hose [4] through a whole [7] of the
lowest the shelf to the water reservoir [8].
3. Prepare a bucket [2].
4. Suck the water, therefore pump the water pump
using the hand grip [6] about four times until
water flows in the outlet hose.
5. Let the water run into the bucket [2] through
gravity.
6. Wipe out the rests of the water at the floor plate of
the water reservoir.
Fig. 17: Water pump