Servicing
Sample/Reference Solenoid Replacement
7-32
Model 49
i
Instruction Manual
Thermo Fisher Scientific
7. Calibrate the instrument. Refer to the “Calibration” chapter in this
manual.
Sample/Reference
Solenoid
Replacement
Use the following procedure to replace a sample/reference solenoid valve.
Equipment required:
Solenoid
Flatblade screwdriver
Open-end wrench, 5/8-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the Sample and Ref connection cables from the
measurement interface board.
3. Remove the plumbing from the solenoids.
4. Loosen the four captive screws and lift out the solenoids and the
solenoid plate.
5. Remove the solenoid retaining nuts from the underside of the faulty
solenoid and remove the solenoid.
6. Replace the new solenoid by following the previous steps in reverse
making sure that all connections are tight.
7. Perform a leak test. Refer to the “Solenoid Leaks” procedure in the
“Preventive Maintenance” chapter.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...