Operation
Service Menu
3-88
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Pressure Calibration
The Pressure Calibration menu is used to calibrate the pressure sensor to
zero, span, or restore factory default values. The pressure calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “Service Mode” earlier in this chapter.
The pressure sensor’s zero counts and span slope are displayed on the menu.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Pressure Calibration
.
Calibrate Pressure Zero
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
•
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
•
Press
to save the current pressure reading as the zero reading.
SINGLE/DUAL/AUTO RANGE:
CURRENTLY: SINGLE
SET TO: DUAL ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
PRESSURE SENSOR CAL:
>ZERO 72
SPAN 1.1447
SET DEFAULTS
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...