Operation
Service Menu
3-96
Model 49
i
Instruction Manual
Thermo Fisher Scientific
•
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Full Scale
. (Hook up a voltage source of 10 V to
the analog input channel.)
•
Use the
and
to increment or decrement the numeric
value.
•
Press
to save the value.
Dilution Ratio
The Dilution Ratio screen allows the user to view and set the dilution ratio.
Acceptable values are 1–500: 1. The default is 1:1. When this value is set,
the dilution ratio is applied to all concentration measurements. This screen
is only accessible if the dilution ratio option is installed.
•
In the Main Menu, choose Service >
Dilution Ratio
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
Display Pixel Test
The Display Pixel Test is used to test the LCD display. The display pixel test
is visible only when the instrument is in service mode. For more
information on the service mode, see “Service Mode” earlier in the chapter.
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 6.24 V
SET TO: 10.0
0
V ?
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
DILUTION RATIO:
CURRENTLY: 001.0 :1
SET TO : 00
2
.0 :1 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
Ü
SAVE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...