System Description
Electronics
Thermo Fisher Scientific
Model 49
i
Instruction Manual
8-7
Flow Sensor Assembly
The flow sensor assembly consists of a board containing an instrumentation
amplifier and a flow transducer with input and output gas fittings. The flow
transducer output is produced by measuring the pressure difference across a
precision orifice. This unit is used for measuring the flow of sample gas in
the measurement system.
Pressure Sensor Assembly
The pressure sensor assembly consists of a board containing an
instrumentation amplifier and a pressure transducer with a gas input fitting.
The pressure transducer output is produced by measuring the pressure
difference between the sample gas pressure and ambient air pressure.
Ozonator Lamp Power Supply
The ozonator lamp power supply board includes the circuits that generate
high voltage AC to control the lamp heater. A transformer outputting high
voltage at approximately 15 kHz drives the lamp. A resistor in series with
the output limits the lamp current. The voltage output is adjustable under
software control for lamp intensity control. The ozonator lamp can be
switched on and off under software control.
The lamp is heated to a specific temperature to ensure reliable operation.
The oscillator circuit that drives the high voltage output is inhibited until
this operating temperature is reached. A power transistor is used to heat the
lamp housing. Thermistors sense the housing temperature for heater control
and for status information.
Photometer Lamp Power Supply
The photometer lamp power supply is similar to the ozonator power supply
except for the operating temperature set point and the value of the resistor
in series with the output. The photometer is always “on” when the
operating temperature is reached. Software on/off control is not required.
Digital Output Board
The digital output board connects to the motherboard and provides
solenoid driver outputs and relay contact outputs to a connector located on
the rear panel of the instrument. Ten relay contacts normally open (with
power off ) are provided which are electrically isolated from each other.
Eight solenoid driver outputs (open collector) are provided along with a
corresp24VDC supply pin on the connector.
I/O Expansion Board
(Optional)
The I/O expansion board connects to the motherboard and adds the
capability to input external analog voltage inputs and to output analog
currents via a connector located on the rear panel of the instrument. It
contains local power supplies, a DC/DC isolator supply, a sub-processor
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...