Operation
Calibration Menu
3-24
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Next Time
The Next Time screen is used to view and set the next zero/span check date
and time. Once the initial zero/span check is performed, the date and time
of the next zero/span check is calculated and displayed.
•
In the Main Menu, choose Calibration > Zero/Span Check >
Next
Time
.
•
Use
,
and
to move and change the value of the date
and time.
•
Press
to edit and accept a change.
Period Hours
The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
•
In the Main Menu, choose Calibration > Zero/Span Check >
Period
Hours
.
ZERO/SPAN CHECK:
>NEXT TIME 01Jan05 12:00
PERIOD HR 24
TOTAL DURATION HR 1.5
ZERO DURATION MIN 30
SPAN DURATION MIN 30
PURGE DURATION MIN 30
ZERO/SPAN AVG SEC 60
ª
RANGE AVG DIAGS ALARM
ZERO CAL RESET
ON
SPAN CAL RESET OFF
ZERO/SPAN RATIO 1
DATE AND TIME:
19 MAR 2005 12:34:56
DATE AND TIME:
PRESS
TO EDIT
19 MAR 2005 12:34:56 ?
SETTING: DAYS
¨
SET MONTHS
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...