Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-71
Set Analog Outputs
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-scale
voltage, zero sets the analog outputs to 0 volts, and normal operation. The
example below shows the selected output state “ALL” is set to normal.
•
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1-6,
or
Current Channel 1-6
.
•
Press
or
to set output.
•
Press
to reset to normal.
Instrument Configuration
The Instrument Configuration screen displays information on the hardware
configuration of the instrument.
Note
If the analyzer is in service mode, pressing
on the item will
toggle it yes or no.
•
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
•
Press
to toggle instrument configuration (in service mode only).
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
SAMPLE/CAL VALVE YES
OZONATOR YES
DILUTION RATIO NO
AUTO CALIBRATION NO
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...