Operation
Instrument Controls Menu
3-40
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Memory Allocation Percent
The Memory Allocation Percent screen is used to select the percentage of
each record type for both short records and long records. Percentages
between 0 and 100% are available in increments of 10. Changing this value
results in log erasure for both short records and long records.
•
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging >
Memory Allocation %
.
•
Use
and
to scroll through a list of choices.
•
Press
to set the percentage for both record types and proceed to
the erasure warning screen.
•
Press
to confirm change.
Data Treatment
The Data Treatment screen is used to select the data type for the selected
record: whether the data should be averaged over the interval, the minimum
or maximum used, or the current value logged. Data treatment doesn’t apply
to all data, just to the concentration measurement. All other data points log
the current value at the end of the interval.
SET PERIOD FOR SREC:
CURRENTLY: OFF MIN
SET TO: 60 MIN ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
SET PERCENT SRECS:
SET PERCENT SRECS:
CURRENTLY: 50 %
CURRENTLY: 50 %
SET TO: 60 % ?
SET TO: 60 % ?
THIS WILL ERASE ALL LOGS!
x
PRESS
¨
TO CONFIRM CHANGE
ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...