Operation
Instrument Controls Menu
3-44
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Streaming Data Interval
The Streaming Data Interval screen is used to adjust the streaming data
interval. The following interval times are available: 1, 2, 5, 10, 20, 30, 60,
90, 120, 180, 240, and 300 seconds.
•
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config >
Streaming Data Interval
.
•
Use
and
to scroll through a list of choices.
•
Press
to save the new streaming data period.
Concentrations
The Concentrations screen allows the user to select the output signal that is
tied to the selected streaming data item. The selected item is shown by “<--”
after it. Range Status is visible only in auto range mode
•
In the Main Menu, choose Instrument Controls > Communication
Settings > Streaming Data Config > Select Item >
Concentrations
.
•
Use
and
to move the cursor up and down.
•
Press
to select a new choice.
STREAMING DATA INTERVAL:
CURRENTLY: 10 SEC
SET TO: 20 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
CONCENTRATIONS:
>NONE
O3 <--
LO O3
HI O3
RANGE STATUS
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...