C-Link Protocol Commands
I/O Configuration
B-42
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
list var aout
list var dout
list var din
These commands report the list of index numbers and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto) for analog output, digital output
and digital inputs. The index number is used to insert the variable in a field
location in a list using “set sp
field index
”. The example that follows reports
the list of analog output, index numbers, and variables.
Send:
list var aout
Receive:
list var aout
index variable
0 none
1 o3
2 no2
5 cellai
6 cellbi
7 noisa
8 noisb
Table B-11.
Default Output Assignment
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
7
Current Output
8
Current Output
9
Current Output
10
Current Output
11
Current Output
12
Current Output
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...