Preventive Maintenance
Safety Precautions
5-2
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Safety Precautions
Read the safety precautions before beginning any procedures in this chapter.
Replacement Parts
See the “Servicing” chapter for a list of replacement parts.
Outside Case
Cleaning
Clean the outside case using a damp cloth being careful not to damage the
labels on the case.
Optical Bench
Cleaning
Best results are obtained when the optical bench is cleaned prior to
recalibration. The cleanliness of the bench should also be checked any time
the detector frequencies drop below 65 kHz, since one source of low output
is light attenuation due to dirt in the cell. Dirt particulates are usually
effective ozone removers.
Use the following procedure to clean the optical bench.
WARNING
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
WARNING
The Model 49
i
Primary Standard is supplied with a three-wire
grounding cord. Under no circumstances should this grounding system be
defeated.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component. For more information
about appropriate safety precautions, see the “Servicing” chapter.
Equipment Damage
Do Not use solvents or other cleaning products to
clean the outside case.
CAUTION
Some internal components can be damaged by small amounts of
static electricity. A properly grounded antistatic wrist strap must be worn
while handling any internal component. For more information about
appropriate safety precautions, see the “Servicing” chapter.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...