Preventive Maintenance
Leak Test and Pump Checkout
5-8
Model 49
i
Instruction Manual
Thermo Fisher Scientific
2. Disconnect the ZERO AIR input lines and plug.
3. If the optional internal ozonator is installed, plug the OZONE and
VENT outputs.
4. Press
to display the Main Menu.
5. Press
to scroll to Diagnostics and press
to display
the Diagnostics menu.
6. Press
to move the cursor to Flows and press
to display the
Flows screen. The flow readings should slowly decrease to zero flow.
7. Press
to return to the Diagnostics menu.
8. Press
to move the cursor to Pressure and press
to display
the Pressure screen. The pressure reading should drop to less than 250
mm Hg.
If the pump diaphragm is in good condition and the capillary is not
blocked, it should take less than 20 seconds from the time the inlet is
plugged to the time the reading below 250 mm Hg is obtained.
If a leak is suspected, carefully tighten each fitting until the leak is found
and check to see that none of the input lines are cracked or broken.
Solenoid Leaks
Leaks across the solenoid valve can be caused by Teflon cold-flowing across
the valve seat or by particulates on the seat.
Use the following procedure to check for solenoid leaks.
1. Generate an ozone concentration of approximately 0.5 ppm.
2. Press
to display the Main Menu.
3. Press
to scroll to Diagnostics and press
to display
the Diagnostics menu.
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...