Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-61
Table Point
The Table Point submenu allows the user to set up an individual table
point.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Point 1-10
.
Volts
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Select Point >
Volts
.
•
Use
and
to move the cursor left or right.
•
Use
and
to move the cursor up and down.
•
Press
to save the new value.
NUMBER OF TABLE POINTS:
CURRENTLY: 2
SET TO: 10 ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
TABLE POINT 01 CONFIG:
>VOLTS 0.00
USER VALUE 0.00
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...