Troubleshooting
Troubleshooting Guides
6-4
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Dirty system
Clean cells and flow
components
Solenoid defective
Perform “Confirmation of
Leak Through Solenoid” test
described in the “Preventive
Maintenance” chapter.
Slow response
Averaging time
Verify averaging time is set
properly
Contaminated optical bench
Clean bench and then
condition system overnight
Table 6-2.
Troubleshooting - Alarm Messages
Alarm Message
Possible Cause
Action
Alarm - O
3
Lamp Temp
Defective lamp heater
Replace ozonator, or lamp
driver board, or lamp
assembly.
Alarm - Lamp Temp
Defective lamp heater
Replace bench lamp driver
board or lamp assembly.
Alarm - Bench Temp
Faulty fan
Replace fan if not operating
properly.
Dirty fan filter
Clean or replace foam filter,
refer to “Preventive
Maintenance” chapter in
this manual.
Alarm - Pressure
High pressure indication
Check the pump for a tear in
the diaphragm, replace with
pump repair kit if necessary.
Refer to “Preventive
Maintenance” chapter in
this manual. Check that
capillaries are properly
installed and O-rings are in
good shape. Replace if
necessary. Check flow
system for leaks.
Alarm - Flow A
Alarm - Flow B
Flow low
Check sample capillary
(0.015 inch ID) for blockage.
Replace as necessary.
Table 6-1.
Troubleshooting - General Guide, continued
Malfunction
Possible Cause
Action
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...