Operation
Range Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-15
information about the dual and auto range modes, see “Single Range
Mode”, “Dual Range Mode”, and “Auto Range Mode” earlier in this
chapter.
Table 3-5 lists the available ranges.
•
In the Main Menu, choose Range >
O
3
Range
.
•
Use
and
to scroll through a list of choices
.
•
Press
to save the new range.
Table 3-5.
Operating Ranges
ppb
ppm
μ
g/m
3
mg/m
3
50
0.05
100
0.1
100
0.10
200
0.2
200
0.20
500
0.5
500
0.50
1,000
1
1,000
1
2,000
2
2,000
2
5,000
5
5,000
5
10,000
10
10,000
10
20,000
20
20,000
20
50,000
50
50,000
50
100,000
100
100,000
100
200,000
200
200,000
200
400,000
400
C1
C1
C1
C1
C2
C2
C2
C2
O3 RANGE:
CURRENTLY: 500.0
SET TO: 1000.0 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...