Operation
Alarms Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-79
Zero and Span Check
The Zero Span Check screen allows the user to view the status of the most
recent zero check and set the maximum zero check offset. The zero and
span check screens are visible only if the zero/span check option is enabled
and function the same way.
•
In the Main Menu, choose Alarms >
Zero
or
Span Check
.
Max Zero and Span Offset
The Max Zero Check Offset screen is used to change the maximum zero
check offset. The maximum zero and span offset screens function the same
way.
•
In the Main Menu, choose Alarms > Zero or Span Check >
Max Offset
.
•
Use
and
to move the cursor left or right.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set to value as actual value.
INTENSITY A:
ACTUAL MIN: 45000 HZ
SET MIN TO: 55000 HZ ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
ZERO CHECK:
ALARM: OK
RESPONSE: 0.0
>MAX OFFSET: 10.0
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...