C-Link Protocol Commands
Keys/Display
B-26
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Send:
set no bkg 5.5
Receive:
set no bkg 5.5 ok
set cal detectors
This command is used to balance the output for both detectors so that they
read the same value
Send:
set cal detectors
Receive:
set cal detectors ok
sp conc
high sp conc
low sp conc
These commands report span concentration in single range mode, or the
high and low span concentrations in dual or auto range mode. If the mode is
incorrect, the instrument responds with “can’t, wrong settings”. The
example below reports the span gas concentration in single range mode.
Send:
sp conc
Receive:
sp conc 1000
set sp conc
value
set high sp conc
value
set low sp conc
value
These commands set the span concentrations to user-defined values to
value
, where
value
is a floating-point representation of the span
concentration in current selected units. The example below sets the span
concentration to 1000 ppb in the single range mode.
Send:
set sp conc 1000
Receive:
set sp conc 1000 ok
Keys/Display
push
button
do
down
en
enter
he
help
le
left
me
menu
ri
right
ru
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...