Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-53
Digital Input Settings
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings
.
•
Use
and
to move the cursor up and down.
•
Press
to select a choice.
Logic State
The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
and closes to trigger the digital input action. If nothing is connected to the
digital input pin, the state should be left at open to prevent the action from
being triggered.
•
Press
to toggle and set the logic state open or closed.
DIGITAL INPUT SETTINGS:
>1 NOP SET BACKGROUND
2 NOP OZONATOR LVL 1
3 NOP OZONATOR LVL 2
4 NOP AOUTS TO ZERO
5 NOP AOUTS TO FS
6 NOP NONE
7 NOP NONE
ª
RANGE AVG DIAGS ALARM
DIGITAL INPUT SETUP:
>LOGIC STATE OPEN
INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...