Operation
Service Menu
3-92
Model 49
i
Instruction Manual
Thermo Fisher Scientific
•
In the Main Menu, choose Service > Select Flow Calibration A or B >
Set Defaults
.
•
Press
to warn user and enable restore with
.
•
Use
to overwrite pressure sensor calibration parameters with
factory default values when pressed after
.
Temperature Calibration
The Temperature calibration screen allows the user to view and set the
ambient temperature sensor calibration. The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see “Service Mode” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Temperature Calibration
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
RESTORE DEFAULT CAL A:
RESTORE DEFAULT CAL A:
RESTORE
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
CALIBRATE AMBIENT TEMP:
CURRENTLY: 32.3
o
C
SET TO: 030.
0
o
C
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...