Servicing
Firmware Updates
7-4
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Figure 7-1.
Properly Grounded Antistatic Wrist Strap
Firmware Updates
The firmware can be updated by the user in the field via the serial port or
over the Ethernet. This includes both the main processor firmware and the
firmware in all low-level processors. Refer to the
iPort
manual for the
firmware update procedure.
Accessing the
Service Mode
If the Service is not displayed on the Main Menu, use the following
procedure to display it.
a. At the Main Menu, press
to scroll to Instrument Controls >
press
>
to scroll to
Service Mode
> and press
.
The Service Mode screen appears.
b. Press
to toggle the Service Mode to ON.
c. Press
>
to return to the Main Menu.
d. Return to the procedure.
Replacement Parts
List
lists the replacement parts for the Model 49
i
major subassemblies.
to identify the component location.
Table 7-1.
Model 49
i
Replacement Parts
Part Number
Description
100480-00
Front Panel Pushbutton Board
101491-15
Processor Board
100533-00
Motherboard
100539-00
Digital Output Board
100542-00
I/O Expansion Board (Optional)
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...