Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 49
i
Instruction Manual
6-3
Lamp
Remove one cell and look for
blue light in hole of input
block
Lamp heater
Check lamp temperature.
From the Diagnostics menu
choose Temperatures.
±15 volt power supply
Check ±15 volts. From the
Diagnostics menu choose
Voltages.
Cell A or B noise
excessive
Foreign material in one cell
Clean cell.
Defective detector
Interchange detectors at
Measurement Interface
Board connectors to
determine if detector is
defective.
Cell A and B noise
excessive
Foreign material in
cells
Clean cells
Lamp failure
Check for 1.7 volt peak to
peak waveform at lamp
current check point on Lamp
Power Supply Board
±15 volt power supply
Check ±15 volts. From the
Diagnostics menu choose
Voltages.
Pressure transducer
does not hold
calibration
Pressure transducer
Replace pressure transducer
Output signal
noisy
Recorder
Replace or repair recorder
Sample is varying
Run instrument on stable
ozone source. If quiet, no
malfunction.
Foreign material in cell
Clean cell
Sticky solenoid valve
Replace with known good
solenoid valve.
Analyzer does not
calibrate properly
Leak
Perform leak test
Contaminated scrubber
Perform scrubber efficiency
test. Replace if necessary.
Pressure transducer out of
calibration
Recalibrate pressure
transducer
Table 6-1.
Troubleshooting - General Guide, continued
Malfunction
Possible Cause
Action
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...