Operation
Calibration Menu
3-22
Model 49
i
Instruction Manual
Thermo Fisher Scientific
In the Main Menu, choose
Calibration
.
Calibrate Zero
The Calibrate Zero screen is used to perform a zero calibration. Be sure the
analyzer samples zero air until the readings stabilize.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more precise the calibration will be. For the most precise
calibration, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “Calibration”.
•
In the Main Menu, choose Calibration >
Calibrate Zero
.
•
Press
to set the new reading to zero.
Calibrate O
3
Coefficient
The Calibrate O
3
Coefficient screen is used to adjust the O
3
coefficient
while sampling span gas of known concentrations. The display shows the
current range. The next line of the display is where the calibration gas
concentration is entered. In dual or auto range modes, “High” or “Low” is
displayed to indicate the calibration of the high or low coefficient.
CALIBRATION:
CALIBRATION:
>CAL ZERO
>CAL ZERO
CAL O3 COEF
CAL O3 HI COEF
ZERO/SPAN CHECK
CAL O3 LO COEF
RANGE AVG DIAGS ALARM
RANGE AVG DIAGS ALARM
O3 BACKGROUND:
O3: 4.4
CURRENTLY: 0.0 ?
SET O3 TO ZERO
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...