Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-33
The Record Display screen (read only) displays the selected records.
•
Use
and
to view all the items in the record.
•
Use
and
to view records at different times.
Date and Time
The Date and Time screen is used to set a start date and time for which to
view logged data. For example, if “20 JUN 2005 10:00” is entered, then the
first logged data record that is displayed is the first record after this time. If
set to one minute logging, this would be at “20 JUN 2005 10:01”.
•
Use
and
to increment or decrement the selected field.
•
Use
to advance to next field.
•
Press
to set the date and time of the first record to be displayed
and continue to the record display screen.
SET # BACK FROM CURRENT:
000000
0
TOTAL LRECS: 20
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
time date flags
10:00 06/20/05 FC0088900
11:00 06/20/05 FC0088900
12:00 06/20/05 FC0088900
13:00 06/20/05 FC0088900
©ª
PGUP/DN
§¨
PAN L/R
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...