C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-13
set alarm conc o3 min
value
set alarm conc o3 max
value
These commands set the O
3
concentration alarm minimum and maximum
value to
value
, where
value
is a floating-point representation of the
concentration alarm limits. The example that follows sets the O
3
concentration alarm maximum value to 150.
Send:
set alarm conc o3 max
Receive:
set alarm conc o3 max 150 ok
alarm intensity a min
alarm intensity a max
alarm intensity b min
alarm intensity b max
These commands report the cell a or b intensity alarm minimum and
maximum value current settings. The example that follows reports that the
cell a minimum value is 45000 Hz.
Send:
alarm intensity a min
Receive:
alarm intensity a min 45000 Hz
set alarm intensity a min
value
set alarm intensity a max
value
set alarm intensity b min
value
set alarm intensity b max
value
These commands set the cell a or b intensity alarm minimum and
maximum value to
value
, where
value
is a representation of the intensity
alarm limits. The example that follows sets the cell a intensity alarm
maximum value to 150000 Hz.
Send:
set alarm intensity a max 150000
Receive:
set alarm intensity a max 150000 ok
alarm level 1
alarm level 2
alarm level 3
alarm level 4
alarm level 5
These commands report the offset value (in ppb) from the concentration
stored at ozonator level 1-5 as the alarm trigger point for that level. The
example that follows reports that the alarm level 1 offset is 20.5.
Send:
alarm level 1
Receive:
alarm level 1 20.5
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...