Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-85
Pump Pressure
The Pump Pressure screen (read only) is used to test the pump. The
solenoids are energized such that there is no flow in Cell B. The flow rate
for Cell B drops to zero and the pressure reading should drop below 390
mmHg in less than 20 seconds. This indicates the effectiveness of the
internal pump. After 20 seconds, if the pressure is greater than 390 mmHg
or the flow is greater than 0.010 LPM, the words “PUMP PROBLEM
DETECTED” appears. After 20 seconds, if the pressure is less than 390
mmHg and the flow is less than 0.010 LPM, the words “PUMP OK”
appears.
•
In the Main Menu, choose Service > Pressure Check >
Pump Pressure
.
Lamp Setting
The Lamp Setting screen is used to adjust the detector intensities. The
display shows the intensities of Cell A and Cell B. The next line of the
display shows the current lamp setting. Adjust the lamp setting until the
intensities are about 100 kHz.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Lamp Setting
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save the new lamp setting.
PUMP PRESSURE: 753.4 mmHg
0.010 LPM
TESTING, WAIT 20 SECS
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...