Setup Mode Menus
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
92
an appropriate scaling factor to the analyzer‟s NO, NO
2
and NO
x
concentration
calculations so that the measurement range and concentration values shown on the
instrument‟s front panel display and reported via the instruments various outputs reflect
the undiluted values.
Using the Dilution Ratio option is a 4-step process:
1. Select the appropriate units of measure (see Section 5.4.3.4).
2. Select the reporting range mode and set the reporting range upper limit (see
Section 5.4.2).
Ensure that the upper span limit entered for the reporting range is the maximum
expected concentration of the
UNDILUTED
gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1
part of sample gas):
Concentration field
displays all gases.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET UNIT
DIL
EXIT
SETUP X.X
O2 DIL FACTOR:1.0 Gain
0
0
0
0
1
.0
ENTR EXIT
Toggle these buttons to
select dilution ratio gain
factor for NO
X
gas.
Default = 1 (e.g. 1:1)
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
NOX DIL FACTOR:1.0 Gain
0
0
0
0
1
.0
ENTR EXIT
Toggle these buttons to select dilution
ratio gain factor for O
2
gas.
NOTE: O
2
Dil Factor only appears if
the optional O
2
Sensor is installed.
4. Calibrate the analyzer.
Ensure that the calibration span gas is either supplied through the same dilution
system as the sample gas or has an appropriately lower actual concentration.
EXAMPLE: If the reporting range limit is set for 100 ppm and the dilution ratio of
the sample gas is 20 gain, either:
a span gas with the concentration of 100 ppm can be used if the span gas
passes through the same dilution steps as the sample gas, or;
a 5 ppm span gas must be used if the span gas
IS NOT
routed through the
dilution system.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...