7.1.3.2. EDITING DAS TRIGGERING EVENTS
Triggering events define when and how the DAS records a measurement of any given
data channel. Triggering events are firmware-specific and a complete list of Triggers for
this model analyzer can be found in Appendix A-5. The most commonly used triggering
events are:
ATIMER
: Sampling at regular intervals specified by an automatic timer. Most
trending information is usually stored at such regular intervals, which can be
instantaneous or averaged.
EXITZR, EXITSP, and SLPCHG (exit zero, exit span, slope change): Sampling at
the end of (irregularly occurring) calibrations or when the response slope changes.
These triggering events create instantaneous data points, e.g., for the new slope
and offset (concentration response) values at the end of a calibration. Zero and
slope values are valuable to monitor response drift and to document when the
instrument was calibrated.
WARNINGS:
Some data may be useful when stored if one of several warning
messages appears such as
WTEMPW
(GFC wheel temperature warning). This is
helpful for troubleshooting by monitoring when a particular warning occurred.
To edit the list of data parameters associated with a specific data channel, follow the
instruction shown in Section 7.1.3 then press:
Starting at the
EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 5, 800
PREV NEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X
NAME: CONC
C
O
N
C
—
—
ENTR EXIT
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
Press each button
repeatedly to
cycle through the available
character set:
0-9, A-Z,
space ’ ~ ! # $ % ^ &
* ( ) - _ = +[ ] { } < >\ | ; : , . / ?
SETUP X.X
NAME: CONC
<SET SET>
EDIT
PRNT
EXIT
Note
A full list of DAS Trigger Events can be found in Appendix A-5 of this manual.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...