Specifications, Approvals, & Compliance
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
20
Parameter
Specification
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs (7 defined, 1 spare)
6 opto-isolated digital control inputs (4 defined, 2 spar)
4 analog outputs
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
Dimensions H x W x D
7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
Weight
Analyzer: 40 lbs (18 kg)
External Pump Pack: 15 lbs (7 kg)
Operating Temperature Range
5 - 40
C (with EPA equivalency)
Humidity Range
0-95% RH non-condensing
Environmental Conditions
Installation Category (Over voltage Category) II Pollution Degree 2
Intended for Indoor Use Only at Altitudes
2000m
1
As defined by the US EPA.
2
Defined as twice the zero noise level by the US EPA.
2.2. EPA REFERENCE DESIGNATION
Teledyne API‟s T200 nitrogen oxides analyzer is designated as an automated reference
method (Number RFNA-1194-099) for NO
2
measurement, as defined in 40 CFR Part
53, when operated under the following conditions:
Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).
Ambient temperature range of 5 to 40°C.
With PTFE filter element or a Kynar® DFU installed in the internal filter assembly.
Equipped with ozone supply air filter
Gas flow supplied by an external vacuum pump capable of ≤10 in-Hg-A at 2
standard liters per minute (slpm).
Software Settings, see Table 2-2:
Table 2-2: T200 Software Settings for EPA Reference
Parameter
Setting
Dynamic Zero
OFF or ON
Dynamic Span
OFF
CAL-on-NO
2
OFF
Dilution Factor
1.0 or OFF
Temp/Pres compensation
ON
AutoCal
ON or OFF
Independent range
ON or OFF
Auto range
ON or OFF
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...