Teledyne API - T200, T204 and 200E Series (05295F DCN6900)
Appendix A3: Warnings and Test Measurements, Software Version K.3
A-31
MODBUS Register
Address
(decimal, 0-based)
Description
10
Units
412
14
O
3
photometer measurement potential
mV
414
14
O
3
cell pressure
PSIA
416
14
O
3
lamp temperature
°C
418
14 + 15
O
3
bench serial number
—
420
14
O
3
bench firmware revision
—
MODBUS Floating Point Holding Registers
(32-bit IEEE 754 format; read/write in high-word, low-word order; read/write)
0 Maps
to
NOX_SPAN1
variable; target conc. for range #1
Conc. units
2 Maps
to
NO_SPAN1
variable; target conc. for range #1
Conc. units
4 Maps
to
NOX_SPAN2
variable; target conc. for range #2
Conc. units
6 Maps
to
NO_SPAN2
variable; target conc. for range #2
Conc. units
100
3
Maps
to
NOX_SPAN3
variable; target conc. for range #3
Conc. units
102
3
Maps
to
NO_SPAN3
variable; target conc. for range #3
Conc. units
200
5
Maps
to
O2_TARG_SPAN_CONC
variable; target conc. for
range O
2
gas
%
300
6
Maps
to
CO2_TARG_SPAN_CONC
variable; target conc. for
range CO
2
gas
%
400
14
Maps to ID_VAR_O3_TARG_SPAN_CONC variable; O
3
target
span concentration
PPB
402
14
Maps to ID_VAR_O3_PRESSURE_OFFSET variable; O
3
cell
pressure compensation offset
PSIA
404
14
Maps to ID_VAR_O3_PRESSURE_SLOPE variable; O
3
cell
pressure slope compensation
—
406
14
Maps to ID_VAR_O3_TEMP_SET variable; O
3
temperature
setpoint
°C
408
14
Maps to ID_VAR_O3_DWELL variable; O
3
dwell time
Seconds
410
14
Maps to ID_VAR_O3_RANGE variable; O
3
analog output
range
PPB
MODBUS Discrete Input Registers
(single-bit; read-only)
0
Manifold temperature warning
1
Converter temperature warning
2 Auto-zero
warning
3
Box temperature warning
4
PMT detector temperature warning
5
Reaction cell temperature warning
6 Sample
flow
warning
7
Ozone flow warning
8
Reaction cell pressure warning
9 HVPS
warning
10
System reset warning
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...