Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
254
12.7.6.4. CONTROL INPUTS
The control input bits can be tested by applying a trigger voltage to an input and
watching changes in the status of the associated function under the
SIGNAL I/O
submenu:
EXAMPLE: to test the “
A
” control input:
1. Under the
DIAG
Signal I/O menu (see Section 12.1.3), scroll through the inputs
and outputs until you get to the output named
EXT_ZERO_CAL
.
2. Connect a jumper from the “+” pin on the appropriate connector to the “U” on the
same connector.
3. Connect a second jumper from the “
” pin on the connector to the “A” pin.
4. The status of EXT_ZERO_CAL should change to read “ON”.
5. Connect a second jumper from the “
” pin on the connector to the “B” pin.
6. The status of EXT_ZERO_CAL should change to read “ON”.
Table 12-11: T200 Control Input Pin Assignments and Corresponding Signal I/O
Functions
INPUT
CORRESPONDING I/O SIGNAL
A
EXT_ZERO_CAL
B
EXT_SPAN_CAL1
C, D, E& F
NOT USED
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...