Communications Setup and Operation
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
126
6.2.2. COM PORT BAUD RATE
To select the baud rate of either COM Port, go to SETUP>MORE>COM and select
either COM1 or COM2 as follows (use COM2 to view/match your personal computer
baud rate when using the USB port, Section 6.6):
EXAMPLE
SETUP X.X
COM1 BAUD RATE:19200
PREV
NEXT
ENTR
EXIT
SETUP X.X
COM1 BAUD RATE:9600
NEXT
ON
ENTR
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X
COM1 BAUD RATE:19200
<SET SET>
EDIT
EXIT
Select which COM
port to configure.
(
COM1
for example).
Use
PREV
and
NEXT
to move between
available baud rates.
300
1200
4800
9600
19200
38400
57600
115200
Press SET> until you
reach the
COM1
BAUD RATE
ENTR
accepts
the new
setting
EXIT
ignores
the new
setting
Figure 6-2.
COM
– COM Port Baud Rate
6.2.3. COM PORT TESTING
The serial ports can be tested for correct connection and output in the COM menu. This
test sends a string of 256 „w‟ characters to the selected COM port. While the test is
running, the red LED on the rear panel of the analyzer should flicker.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...