123
6.
COMMUNICATIONS SETUP AND OPERATION
This instrument‟s rear panel connections include an Ethernet port, a USB port (option)
and two serial communications ports labeled
RS232
, which is the COM1 port, and
COM2
(refer to Figure 3-4). These ports give the user the ability to communicate with,
issue commands to, and receive data from the analyzer through an external computer
system or terminal. Connection instructions were provided in Section
This section provides pertinent information regarding communication equipment,
describes the instrument‟s communications modes, presents configuration instructions
for the communications ports, and provides instructions for their use, including
communications protocol. Data acquisition is presented in Section
6.1. DATE TERMINAL / COMMUNICATION EQUIPMENT
(DTE DCE)
RS-232 was developed for allowing communications between data terminal equipment
(DTE) and data communication equipment (DCE). Basic terminals always fall into the
DTE category whereas modems are always considered DCE devices. The difference
between the two is the pin assignment of the Data Receive and Data Transmit functions.
DTE devices receive data on pin 2 and transmit data on pin 3.
DCE devices receive data on pin 3 and transmit data on pin 2.
To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers
(which can be either), a switch mounted below the serial ports on the rear panel labeled
DCE
DTE
(Figure 3-4) allows the user to set the RS-232 configuration for one of these
two data devices. This switch exchanges the Receive and Transmit lines on RS-232
emulating a cross-over or null-modem cable. The switch has no effect on COM2.
6.2. COMMUNICATION MODES, BAUD RATE AND PORT TESTING
Use the SETUP>MORE>COM menu to configure COM1 (labeled
RS232
on instrument
rear panel) and/or COM2 (labeled
COM2
on instrument rear panel) for communication
modes, baud rate and/or port testing for correct connection. If using a USB option
communication connection, setup requires configuring the COM2 baud rate (Section
6.2.2).
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...