Analyzer Operation Manual
Getting Started
69
3.4.4.2. INITIAL CALIBRATION PROCEDURE FOR T200 ANALYZERS WITHOUT OPTIONS
The following procedure assumes that:
The instrument DOES NOT have any of the available calibration valve or gas inlet
options installed;
Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer
and;
The pneumatic setup matches that described in Section 3.3.2.
V
ERIFYING THE
R
EPORTING
R
ANGE
S
ETTINGS
Although you can use any range setting, we recommend that you perform this initial
checkout using following reporting range settings:
Unit of Measure: PPB
Reporting Range: 500 ppb
Mode Setting: SNGL
While these are the default settings for the T200 analyzer, it is recommended that you
verify them before starting the calibration procedure, by pressing the following menu
button sequence: