Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Getting Started
71
To verify that the analyzer span setting is set for
400 PPB
, press:
SAMPLE RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL
SETUP
M-P CAL RANGE=500.0 PPB
NOX=XXXX
<TST TST>
ZERO SPAN
CONC
EXIT
M-P CAL CONCENTRATION MENU
NOX
NO
CONV
EXIT
M-P CAL
[GAS TYPE]
SPAN CONC:400.0 Conc
0
4
0
0
.0
ENTR EXIT
The NO
X
& NO span concentration
values automatically default to
400.0 PPB
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the NO
X
and NO
calibration gases.
EXIT
ignores the new
setting and returns to
the previous display.
ENTR
accepts the new
setting and returns to the
CONCENTRATION MENU.
If using NO span gas in
addition to NO
X
repeat last
step.
I
NITIAL
Z
ERO
/S
PAN
C
ALIBRATION
P
ROCEDURE
To perform an initial calibration, press:
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...