
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
Setup Mode Menus
111
5.9.3.6. MANUAL CALIBRATION OF THE ANALOG OUTPUTS
CONFIGURED FOR VOLTAGE
RANGES
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Note
The menu for manually adjusting the analog output signal level will only appear
if the AUTO-CAL feature is turned off for the channel being adjusted
. (See
Section 5.9.3.3).
Calibration is performed with a voltmeter connected across the output terminals and by
changing the actual output signal level using the front panel buttons in 100, 10 or 1
count increments. See Figure 3-8 for pin assignments and diagram of the analog output
connector.
V
+DC Gnd
Recording
Device
V IN +
V IN -
ANALYZER
V OUT +
V OUT -
Volt
Meter
Figure 5-6:
Setup for Checking / Calibrating DCV Analog Output Signal Levels
Table 5-7: Voltage Tolerances for the TEST CHANNEL Calibration
FULL
SCALE
ZERO
TOLERANCE
SPAN VOLTAGE
SPAN
TOLERANCE
MINIMUM
ADJUSTMENT
(1 count)
0.1 VDC
±0.0005V
90 mV
±0.001V
0.02 mV
1 VDC
±0.001V
900 mV
±0.001V
0.24 mV
5 VDC
±0.002V
4500 mV
±0.003V
1.22 mV
10 VDC
±0.004V
4500 mV
±0.006V
2.44 mV
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...