5.4.3.3. SETUP
RNGE
MODE
AUTO: CONFIGURING THE T200 ANALYZER FOR AUTO
RANGE MODE
In AUTO range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (LOW and HIGH). The same low and high span settings are
applied equally to NO, NO
2
and NO
X
readings.
The unit will switch from
LOW
range to
HIGH
range when either the NO, or NO
X
concentration exceeds 98% of the low range span.
The unit will return from
HIGH
range back to
LOW
range once both the NO and
NO
X
concentrations fall below 75% of the low range span.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
The LOW & HIGH ranges referred to here are NOT the same as the low & high
physical ranges referred to in Section 5.4.1.
Also the
RANGE
test function displayed on the front panel will be replaced by two
separate functions:
RANGE1: The LOW range setting for all analog outputs
RANGE2: The HIGH range setting for all analog outputs
The
LOW/HIGH
range status is also reported through the external, digital status bits
(Section 3.3.1.4).
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...