Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
232
12.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS
In addition to being useful as predictive diagnostic tools, the test functions viewable
from the analyzers front panel can be used to isolate and identify many operational
problems when combined with a thorough understanding of the analyzer‟s principles of
operation (see Section 13).
The acceptable ranges for these test functions are listed in the “Nominal Range” column
of the analyzer
Final Test and Validation Data Sheet
(P/N 04490) shipped with the
instrument. Values outside these acceptable ranges indicate a failure of one or more of
the analyzer‟s subsystems. Functions whose values are still within acceptable ranges but
have significantly changed from the measurement recorded on the factory data sheet
may also indicate a failure.
A worksheet has been provided in Appendix C to assist in recording the value of these
test functions.
Note
A value of ―XXXX‖ displayed for any of these TEST functions indicates an OUT
OF RANGE reading.
Note
Sample Pressure measurements are represented in terms of ABSOLUTE pressure
because this is the least ambiguous method reporting gas pressure.
Absolute atmospheric pressure is about 29.92 in-Hg-A at sea level. It decreases
about 1 in-Hg per 1000 ft gain in altitude. A variety of factors such as air
conditioning systems, passing storms, and air temperature, can also cause
changes in the absolute atmospheric pressure.
Table 12-2: Test Functions - Indicated Failures
TEST FUNCTION
INDICATED FAILURE(S)
NOX STB
Unstable concentrations; leaks
SAMP FlW
Leaks; clogged critical flow orifice
OZONE FL
Leaks; clogged critical flow orifice
PMT
Calibration off; HVPS problem; no flow (leaks)
NORM PMT
Auto Zero too high
AZERO
Leaks; malfunctioning NO, NO
x
or Auto Zero valve; O
3
air filter cartridge exhausted
HVPS
Calibration off; preamp board circuit problems
RCELL TEMP
Malfunctioning heater; relay board communication (I
2
C bus); relay burnt out
BOX TEMP
Environment out of temperature operating range; broken thermistor
PMT TEMP
TEC cooling circuit broken; relay board communication (I
2
C bus); 12 V power supply
IZS TEMP (option)
Malfunctioning heater; relay board communication (I
2
C bus); relay burnt out
MOLY TEMP
Malfunctioning heater; disconnected or broken thermocouple; relay board communication
(I
2
C bus); relay burnt out; incorrect AC voltage configuration
RCEL (pressure)
Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices
SAMP (pressure)
Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices; sample inlet
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...