Calibration Procedures
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
196
9.4.3.2. ZERO/SPAN POINT CALIBRATION PROCEDURE
Press ENTR to changes
the OFFSET & SLOPE
values for both the NO
and NO
x
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
NOX STB= XXX.X PPB
NOX=XXX.X
< TST TST > CAL SETUP
M-P CAL
NOX STB= XXX.X PPB
NOX=XXX.X
<TST TST>
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes
the OFFSET & SLOPE
values for both the NO
and NO
x
measurements.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
Set the Display to show
the NOX STB test
function.
This function calculates
the stability of the NO/NO
x
measurement
Toggle TST> button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST> ZERO
SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until NOX STB
falls below 0.5 PPB.
This may take several
minutes.
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
NOX STB= XXX.X PPB
NOX=X.XXX
<TST TST> ENTR
CONC
EXIT
The SPAN button now
appears during the transition
from zero to span.
You may see both buttons.
If either the ZERO or SPAN
buttons fail to appear see
Section 11 for
troubleshooting tips.
Analyzer continues to
cycle through NO
x
,
NO, and NO
2
measurements
throughout this
procedure.
Only appears if the AUTO
range mode is selected.
Press to select the
appropriate range.
Repeat entire procedure for
each range.
Only appears if the AUTO
range mode is selected.
Press to select the
appropriate range.
Repeat entire procedure for
each range.
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST
TST>
CAL CALZ CALS
SETUP
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST TST> CAL
CALZ
CALS
SETUP
SAMPLE
RANGE=500.0 PPB NOX= XXXX
<TST TST> CAL CALZ
CALS
SETUP
Note
If the ZERO or SPAN buttons are not displayed, the measurement made during is
out of the allowable range allowed for a reliable calibration. See Section 12 for
troubleshooting tips.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...