Troubleshooting & Service
Teledyne API
– T200 NO/NO2/NOx Analyzer Operation Manual
242
12.5. CALIBRATION PROBLEMS
This section describes possible causes of calibration problems.
12.5.1. NEGATIVE CONCENTRATIONS
Negative concentration values can be caused for several reasons:
A slight, negative signal is normal when the analyzer is operating under zero gas
and the signal is drifting around the zero calibration point.
This is caused by the analyzer’s zero noise and may cause reported
concentrations to be negative for a few seconds at a time down to -20 ppb, but
should randomly alternate with similarly high, positive values.
The T200 has a built-in Auto Zero function, which should take care of most of
these deviations from zero, but may yield a small, residual, negative value.
If larger, negative values persist continuously, check if the Auto Zero function
was accidentally turned off using the remote variables in Appendix A-2.
In this case, the sensitivity of the analyzer may be drifting negative.
A corruption of the Auto Zero filter may also cause negative concentrations.
If a short, high noise value was detected during the Auto Zero cycle, that higher
reading will alter the Auto Zero filter value.
As the value of the Auto Zero filter is subtracted from the current PMT
response, it will produce a negative concentration reading.
High Auto Zero readings can be caused by
a leaking or stuck Auto Zero valve (replace the valve),
by an electronic fault in the preamplifier causing it to have a voltage on the
PMT output pin during the Auto Zero cycle (replace the preamplifier),
by a reaction cell contamination causing high background (>40 mV) PMT
readings (clean the reaction cell),
by a broken PMT temperature control circuit, allowing high zero offset
(repair the faulty PMT cooler). After fixing the cause of a high Auto Zero
filter reading, the T200 will take 15 minutes for the filter to clear itself, or
by an exhausted chemical in the ozone cleanser (see Section 11.3.3).
Calibration error is the most likely explanation for negative concentration values.
If the zero air contained some NO or NO
2
gas (contaminated zero air or a worn-
out zero air scrubber) and the analyzer was calibrated to that concentration as
“zero”, the analyzer may report negative values when measuring air that
contains little or no NO
x
.
The same problem occurs, if the analyzer was zero-calibrated using zero gas
that is contaminated with ambient air or span gas (cross-port leaks or leaks in
supply tubing or user not waiting long enough to flush pneumatic systems).
If the response offset test functions for NO (
NO OFFS
) or NO
X
(
NOX OFFS
) are
greater than 150 mV, a reaction cell contamination is indicated.
Clean the reaction cell as described in Section 11.3.9.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...