Teledyne API - T200, T204 and 200E Series (05295F DCN6900)
APPENDIX A-2: Setup Variables
A-9
APPENDIX A-2: Setup Variables
Table A-1: Setup Variables
Setup Variable
Numeric
Units
Default
Value
Value Range
Description
DAS_HOLD_OFF
Minutes
15
0.5–20
Duration of DAS hold off period.
MEASURE_MODE —
NO-NOX,
NOX
4
NO,
NOX, NOX-NO,
NON-OX
Gas measure mode. Enclose value in
double quotes (") when setting from the
RS-232 interface.
STABIL_GAS — NOX
NO,
NO2,
NOX,
O2
5
,
CO2
6
Selects gas for stability measurement.
Enclose value in double quotes (") when
setting from the RS-232 interface.
TPC_ENABLE — ON OFF, ON
ON enables temperature/ pressure
compensation; OFF disables it.
DYN_ZERO —
OFF
ON, OFF
ON enables remote dynamic zero
calibration; OFF disables it.
DYN_SPAN —
OFF
ON, OFF
ON enables remote dynamic span
calibration; OFF disables it.
IZS_SET
1
ºC
51
30–70
IZS temperature set point and warning
limits.
Warnings:
50–52
CONC_PRECISION —
AUTO
1
,
3
2, 3
AUTO,
0,
1,
2,
3,
4
Number of digits to display to the right of
the decimal point for concentrations on the
display. Enclose value in double quotes (")
when setting from the RS-232 interface.
STAT_REP_GAS
4
—
NOX
NO,
NO2,
NOX,
CO2
6
,
O2
5
Selects gas to report in TAI protocol status
message. Enclose value in double quotes
(") when setting from the RS-232 interface.
REM_CAL_DURATI
ON
4
Minutes 20
1–120
Duration of automatic calibration initiated
from TAI protocol.
CLOCK_ADJ
Sec./Da
y
0
-60–60
Time-of-day clock speed adjustment.
CAL_ON_NO2
1
— OFF ON,
OFF
ON enables span calibration on pure NO
2
;
OFF disables it.
SERVICE_CLEAR —
OFF
OFF
ON
ON resets the service interval timer.
TIME_SINCE_SVC Hours
0
0–500000
Time since last service.
SVC_INTERVAL
Hours
0
0–100000
Sets the interval between service
reminders.
1
T200 and M200E.
2
T200H and M200EH.
3
T200U and M200EU.
4
TAI
protocol
5
O
2
option.
6
CO
2
option.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...