Data Acquisition System (DAS) and APICOM
Teledyne API
– T200 NO/NO
2
/NO
X
Analyzer Operation Manual
158
To modify, add or delete a parameter, follow the instruction shown in Section 7.1.3 then
press:
Starting at the
EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 5, 800
PREV NEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X
NAME: CONC
<SET
SET>
EDIT
EXIT
Continue pressing
<SET
or
SET>
until ...
SETUP X.X
PARAMETER:5
<SET SET
>
EDIT
EXIT
SETUP X.X
0) PARAM=NXCNC1, MODE=AVG
PREV NEXT
INS
DEL
EDIT
EXIT
SETUP X.X
EDIT PARAMS (DELETE DATA)?
YES
NO
YES
deletes all data
currently stored for
this data channel and
continues into
EDIT
mode.
NO
retains the
data and
returns to the
previous
menu.
SETUP X.X
PARAMETER:NXCNC1
<SET
SET> EDIT
EXIT
SETUP X.X
PARAMETER:NXCNC1
PREV NEXT
ENTR EXIT
SETUP X.X
SAMPLE MODE:AVG
<SET
SET> EDIT
EXIT
SETUP X.X
PARAMETER:NXCNC1
INST AVG SDEV MIN
MAX
ENTR EXIT
Press the desired
MODE
button.
SETUP X.X
PRECISION:4
<SET
SET> EDIT
EXIT
SETUP X.X
PRECISION:5
5
ENTR EXIT
Toggle this button to
set from 1 to 4.
SETUP X.X
STOR NUM SAMPLE:OFF
<SET
EDIT
EXIT
SETUP X.X
STOR NUM SAMPLE:OFF
OFF
ENTR EXIT
Toggle this button to
turn
ON/OFF.
Pressing
<SET
returns to the
previous Function.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
Toggle these buttons to
cycle through the list of
available parameters.
PREV
INS
NEXT
DEL
Button
Exports the configuration of all data channels to the
RS-232 interface
Selects the previous data channel or parameter
Inserts a new data channel or parameter into the list
BEFORE the selected channel
Selects the next data channel or parameter
Deletes the currently selected data channel or
parameter
FUNCTION
Buttons only appear when applicable
DAS EDIT
– Touchscreen Functions
EDIT
Enters EDIT mode
<SET
SET>
Selects the previous property to be edited
Selects the next property to be edited
Toggle these
buttons to select a
different parameter.
Note
When the STORE NUM SAMPLES feature is turned on, the instrument will store
the number of measurements that were used to compute the AVG, SDEV, MIN or
MAX value but not the actual measurements themselves.
06858E DCN7057
Содержание T200
Страница 2: ......
Страница 4: ...ii This page intentionally left blank 06858E DCN7057 ...
Страница 8: ...vi This page intentionally left blank 06858E DCN7057 ...
Страница 344: ...Index Teledyne API T200 NO NO2 NOX Analyzer Operation Manual 342 This page intentionally left blank 06858E DCN7057 ...
Страница 380: ...This page intentionally left blank B 2 06858E DCN7057 ...
Страница 384: ...This page intentionally left blank B 6 06858E DCN7057 ...
Страница 387: ...APPENDIX D Wire List and Electronic Schematics 06858E DCN7057 D 1 ...
Страница 388: ...This page intentionally left blank D 2 06858E DCN7057 ...
Страница 393: ...06858E DCN7057 D 7 ...